ZR-7250 Air Quality Monitoring Station

Short Description:

Unlike other sensor-based instruments, the ZR-7250 Air Quality Monitoring Station is designed to be calibrated using standard calibration equipment used to calibrate ambient air quality analyzers. This ensures your measurements will be robust and traceable back to reference standards. We also offer calibration equipment specifically designed for the ZR-7250  the ZR-5409  portable calibrator and the  ZR-5409 which comes fully integrated with your ZR-7250 system.


  • CO Range: (0~50)umol/mol
  • SO2 Range: (0~500)umol/mol
  • NOx Range: (0~500)nmol/mol
  • O3 Range: (0~500)nmol/mol
  • PM10/PM2.5/PM1 Range: (0~1000)μg/m3 or(0~10000)μg /m3
  • Product Detail

    Specification

    Air Quality Monitoring Station (AQMS) is a system that measures metrological parameters such as temperature, humidity, barometric pressure, wind speed, wind direction, noise and ambient parameters. The AQMS also integrates a series of ambient analyzers to monitor the concentration of air pollutants (such as SO2, NOX, CO, O3, PM10, PM2.5 etc.) real-time and continuously.

    Suitable for use on a range of projects, including national and urban air monitoring networks, roadside monitoring, and industrial perimeter monitoring.

    Who is ZR-7250 for?

    Researchers, air monitoring professionals, environmental consultants, and industrial hygienists use the ZR-7250 AQMS to set up national and urban air monitoring networks, conduct environmental impact assessments, and to ensure sensitive receptors in the community are not at risk from air pollution.

     

    What can ZR-7250 measure?

     Particulate Matter: PM10, PM2.5, PM1

    Gases: SO2, NOX, CO, O3

    Environmental: Temperature, humidity, noise, barometric pressure, wind speed and direction

    Practical applications for the ZR-7250 AQMS include:

     Urban air monitoring networks

    >  National air monitoring networks

    >  Roadside air monitoring

     Industrial perimeter monitoring

     

     Environmental impact assessments

     Research and consultancy projects

     Short term hot spot monitoring

    Features

     Continuous, simultaneous measurement of up to 10 common air pollutants and environmental parameters in real-time.

     The AQMS Series can be customized. A unique modular design increases flexibility and makes maintenance and servicing easy.

     

    The station can also be equipped with integrated calibration.

    Data is traceable back to international standards – USEPA (40 CFR Part 53) and EU (2008/50/EC).

    Remote data transmission, powerful data storage function up to one year.

    1

     

     

    Deliver Goods

    deliver goods Italy
  • Previous:
  • Next:

  • Parameter

    CO

    SO2

    NOx

    O3

    Principle

    NDIR

    UV Fluorescence

    CLIA

    UV Spectrophotometry

    Range

    (0~50)umol/mol

    (0~500)umol/mol

    (0~500)nmol/mol

    (0~500)nmol/mol

    Sampling flowrate

    (800-1500)mL/min

    (500-1000)mL/min

    (450±45)mL/min

    800 mL/min

    The lowest detect limit

    ≤0.5 umol/mol

    ≤2 umol/mol

    ≤0.5 nmol/mol

    ≤1 nmol/mol

    Error

    ±2%F.S.

    ±5%F.S.

    ±3%F.S.

    ±2%F.S.

    Response

    ≤4min

    ≤5min

    ≤120s

    ≤30s

    Data storage

    250000 groups

    Size

    (L494*W660*H188)mm

    Weight

    15kg

    Power supply

    AC(220±22)V,(50±1)Hz

    Consumption

    ≤300W

    ≤300W

    ≤700W

    ≤300W

     

    Parameter

    PM10/PM2.5/PM1

    Principle

    Beta Attenuation method

    Range

    (0~1000)μg/m or(0~10000)μg /m3

    Sampling flowrate

    16.7L/min

    Sampling cycle

    60min

    Atmospheric pressure

    (60~130)kPa

    Humidity

    (0~100)%RH

    Data storage

    365 day hourly concentration data

    Size

    (L324*W227*H390)mm

    Weight

    11kg(Sampling head included)

    Consumption

    ≤150W

    Power supply

    AC(220±22)V,(50±1)Hz